• Sourcing Solutions
  • Services & Membership
  • Help & Community

Quick Details

  • Place of Origin: Beijing, China (Mainland)
  • Brand Name: GFUVE
  • Model Number: GF302C
  • Power: Electronic
  • Usage: Universal Testing Machine
  • GF302C: Portable Multi-function Instrument Calibrator
  • AC Voltage signal output: 50V--800V
  • Certification: ISO 9001 : 2008, CE
  • AC current signal output: 0.5A--20A
  • Voltage / current harmonic output: 2--31st
  • Operating temperature: 0 ~ 40°
  • Dimensions: 460 mm x 430 mm x 185 mm
  • Frequency: 45--65HZ
  • Adjust fineness: 5 x 10-5
  • Warranty: 2 years

Packaging & Delivery

Packaging Details Inside a layer of aluminum alloy box, outside is packed in wooden cases
Delivery Time Shipped in 15 days after payment

GF302C multifunction instrument test equipment

 

1. Function: Measure Voltage, Current, Vector, Waveform Distortion, Harmonic,Power, Phase, Power Factor, Frequency, Energy Error.

2.Wide range: Voltage 60V-480V, Current 1A,10A, Auto switching range/Manual switching range.

3.Complete software calibration, to eliminate the drawback of hardware calibration.

Specifications

 

 

AC voltage output 

AC current output 

DC voltage output 

DC current output 

Range

50V---800V

0.5A---20A

75V--1000V

0.5A---20A

Adjusting range

 (0-120) % RG 

(0-120) % RG 

 (0-120) % RG 

 (0-120) % RG 

Adjust fineness

0.01% RG 

0.01% RG 

0.005% RG

0.005% RG

Accuracy

0.05% RG 

0.05% RG 

0.05% RG

0.05% RG

Stability

0.01% / 1 min 

0.01% / 1 min 

 0.01% / 1 min 

0.01% / 1 min 

Load capacity

25 VA 

 25 VA 

 25 VA 

 25 VA 

Output distortion degree

≤0.3% or (linear load) 

≤0.3% or (linear load) 

  

 

AC power output 

Accuracy: 0.05% RG     Stability: 0.01% / 1 min 

Frequency 

Frequency range: 45.000 ~ 65.000 Hz    Resolution: 0.001 Hz    Accuracy: 0.01 Hz 

Power factor output

Adjusting range: -1~0~1     Adjust fineness: 0.000     Accuracy: 0.0005

Phase angle

Scope: 0.00 ° ~ 359.99 °   Resolution: 0.01 °    Accuracy: 0.05 ° 

Voltage/Current harmonic output

2~31st

DC measurements 

DC voltage measurement range: 0 ~ ± 24V      DC current measurement range: 0 ~ ± 24mA 

Measurement accuracy: 0.01% RG 

Watt-hour meter measuring the integrated error 

Active power: 0.05%       Reactive power: 0.1% 

Environmental conditions

Working temperature: 0 ~ 40°    Relative humidity:  85%    Storage conditions: -30°C ~ 60°C 

Work power 

Single phase AC 220 V±10%, Frequency 50/60 Hz 

Volume and weight 

Size: 460 mm x 430 mm x 185 mm 

Weight: about 20 Kg 

Our Service

We are able to provide Complete set of scan head according to your detailed requirements,please contact me directly by skype or send inquiries to me at any time!we are always in a position to quote you the most advantageous price for higher quality merchandise . ^_^ 

 

 

Certifications

 

Transaction History of the Supplier

Below is the information about the supplier's transactions conducted via Alibaba.com. If you require further details regarding the transaction data, please contact the supplier directly.

Transaction Overview

Transactions

Transaction Details

This supplier has completedtransactions with buyers from .

Shipping Destination
Transaction Value
Transaction Date
Email to this supplier
China (Mainland) | Trading Company
Supplier Assessments:
The supplier supports Trade Assurance – A free service that protects your orders from payment to delivery.
Supplier's Trade Assurance Limit: US $49,000
Learn More